Electrospray Ionisation and Why Peptides Show Multiple Charge States

One peptide, several peaks. This is what electrospray ionisation does, why it is expected rather than alarming, and what a matching mass does and does not establish.

Novagen Analytical Labs

Electrospray ionisation (ESI) turns a peptide dissolved in liquid into charged molecules in the gas phase without breaking the peptide apart. It does this by spraying the solution through a charged needle, so the peptide leaves as ions that carry one or more protons. Because the number of protons varies from molecule to molecule, a single compound appears as several peaks at different m/z values. Software then collapses those peaks back to one neutral mass.

That is the short answer. The detail matters because the multiple peaks are the first thing a reader sees on a spectrum, and they are easily misread as a mixture. They are not. They are the normal signature of a peptide measured by ESI.

How electrospray ionisation works

The sample is dissolved in a mixture of water and an organic solvent, usually with a little acid. The solution flows through a fine capillary held at a high voltage relative to the instrument inlet. The strong electric field acts on the liquid at the tip.

The liquid at the tip is drawn into a cone shape, called a Taylor cone. A fine jet leaves the cone and breaks into small droplets. Each droplet carries an excess of charge of one sign, positive in positive ion mode. A stream of warm gas assists the spray and speeds up evaporation.

As solvent evaporates, each droplet shrinks and its charge stays put. The charge per unit of surface area rises. At a point known as the Rayleigh limit, the repulsion between the charges matches the surface tension holding the droplet together. The droplet then releases a set of smaller, highly charged offspring droplets. This is Coulombic fission. The cycle repeats until the droplets are very small.

The final step, in which a single gas-phase ion is released, is described by two models. In the ion evaporation model, an ion is ejected from the surface of a very small droplet by the field. In the charged residue model, the droplet evaporates completely and leaves its charge on the analyte. The literature generally associates the first with small ions and the second with large, folded macromolecules. Direct experimental evidence to settle the question for every case is limited. For peptides, the practical point is the same under either model: the molecule arrives in the gas phase carrying charge.

Why ESI is called a soft technique

A soft ionisation method adds little internal energy to the molecule. Some other approaches deposit enough energy to break bonds, and the spectrum then shows fragments instead of the whole molecule. That can be useful for some purposes. It is a problem when the aim is to confirm the mass of the intact peptide.

In ESI the peptide never meets a high-energy electron beam. It is carried in droplets, and its charge comes from protons transferred in solution or at the droplet surface. Covalent bonds generally survive. The molecular ion is the main thing observed.

Soft does not mean gentle under all conditions. Source voltages and temperatures can be set high enough to cause in-source fragmentation or to lose water or ammonia from a molecule. A well-run method keeps the settings mild. A spectrum is evidence about the sample and the method together.

The m/z relationship

A mass spectrometer does not measure mass. It measures mass divided by charge, written m/z. For a peptide with neutral monoisotopic mass M that has gained z protons, the ion appears at:

m/z = (M + z x 1.00728) / z

The value 1.00728 is the mass of a proton. It is used instead of the hydrogen atom mass (1.00783) because the added particle is a bare proton and the electron stays with the atom it came from.

Take an illustrative peptide with M = 4000.00 Da. It is a round number chosen for clarity and is not a specific product.

  • z = 3: (4000.00 + 3 x 1.00728) / 3 = 4003.02184 / 3 = 1334.341
  • z = 4: (4000.00 + 4 x 1.00728) / 4 = 4004.02912 / 4 = 1001.007
  • z = 5: (4000.00 + 5 x 1.00728) / 5 = 4005.03640 / 5 = 801.007

One molecule, three peaks, at m/z 1334.341, 1001.007 and 801.007. A reader who sees only the x-axis could believe three different compounds are present. The arithmetic shows they are one.

Why multiple charge states appear

A peptide has several sites that accept a proton. The N-terminal amine is one. The side chains of lysine, arginine and histidine are others, and arginine is the most basic of the group. In acidic solution many of these sites are protonated.

Not every molecule in the spray ends up with the same number of protons. The outcome depends on the droplet it was in, how fast the droplet dried and how the charge was shared. The result is a distribution. Some molecules carry three charges, more carry four, some carry five. On the spectrum this looks like a bell-shaped envelope of peaks, with the most intense peak at the most common charge state.

Two general patterns are reliable:

  • Larger peptides tend to carry more charges, because they have more basic sites and more surface to hold charge.
  • Peptides with more basic residues tend to show higher charge states than peptides of similar size with fewer.

Solvent composition and acid content also shift the envelope. A change in the charge distribution between two runs does not by itself indicate a change in the sample.

Charge deconvolution: from many peaks to one mass

Deconvolution is the step that converts the m/z series into a neutral mass. The reasoning can be done by hand with two adjacent peaks.

Take the z = 4 peak at 1001.007 and the z = 5 peak at 801.007. Call the higher one m1 and the lower m2. For adjacent charge states, the higher peak's charge is:

z = (m2 - 1.00728) / (m1 - m2)

Here that is (801.007 - 1.007) / (1001.007 - 801.007) = 800 / 200 = 4. With z known, the neutral mass follows directly: M = 4 x (1001.007 - 1.007) = 4000.00 Da. The other peaks give the same answer, which is a useful internal check. If the calculated masses from different peaks disagree, something other than a simple charge series is present.

Instrument software does the same thing across the whole envelope, with statistical fitting. It reports either a monoisotopic mass or an average mass. The difference matters. The monoisotopic mass uses the lightest isotope of each element and is the number to use for a small peptide on a high-resolution instrument. The average mass is weighted by natural isotope abundance and is closer to what a large molecule or a low-resolution instrument reports. A certificate should say which one it is quoting.

Reading charge state from isotope spacing

A peptide contains carbon, and about 1.1 percent of carbon is the heavier isotope carbon-13. A molecule therefore appears as a cluster of peaks, one for each count of heavy atoms. The clusters are separated by about 1.003 Da in mass.

On the m/z axis, that spacing is divided by the charge. The isotope peaks of a singly charged ion are about 1.003 apart. A doubly charged ion has peaks about 0.5 apart. For the z = 4 ion above, the spacing is about 0.25, and for z = 5 it is about 0.20.

This allows the charge state to be read from a single cluster, without needing neighbouring peaks. It requires enough resolving power to separate the isotope peaks. On a low-resolution instrument the cluster merges into one broad peak and the charge must be found from the series instead.

Polarity and adducts

Peptides are usually run in positive ion mode, where the ions are [M+nH]n+. The basic sites accept protons readily. Negative ion mode forms [M-nH]n- by removing protons, with m/z = (M - n x 1.00728) / n. It is used for peptides that are rich in acidic residues, or for modified molecules, but it is the less common choice.

Ions can also form by picking up something other than a proton. These are adducts. The common ones in positive mode come from sodium, potassium and ammonium. Relative to the protonated ion, a sodium adduct sits about 21.98 Da higher, a potassium adduct about 37.96 Da higher, and an ammonium adduct about 17.03 Da higher, all divided by the charge on the m/z axis.

Sodium and potassium come from glassware, solvents and the sample itself. Adducts are a normal part of a spectrum. They split intensity between more peaks and can make a deconvolution look untidy, but they carry the same information about the molecule. A reader who knows these mass offsets can recognise them rather than mistake them for impurities.

What this means on a certificate

On a certificate, the ESI result is an identity check. The laboratory reports an observed mass, derived by deconvolution, next to the expected mass from the declared structure. The two agree within a stated tolerance, or they do not.

A match tells you that a species with the expected mass is present in the sample submitted. That is useful and it is limited. It does not establish sequence, because different sequences can have identical mass. See molecular weight and identity for the detail. It does not measure purity, because ESI intensity depends on how readily each compound ionises, and peaks are not proportional to amounts. Purity comes from a separation method, as covered in HPLC versus mass spectrometry and UV detection wavelength. And purity is composition, not safety.

A mass that does not match is also informative. It can mean a different compound, a modification, a truncation, or a salt or adduct that was not accounted for in the calculation.

Reading a report

When you read an ESI result, four checks are worth making. First, confirm that the charge states listed give consistent neutral masses. Second, check whether the quoted mass is monoisotopic or average. Third, compare the observed mass with the expected one and note the tolerance. Fourth, look for unexplained peaks and ask whether adducts or fragments account for them.

Novagen is a third-party analytical laboratory. It reports what was measured in the sample submitted, and nothing beyond it. Certificates can be checked against the published register at the verification page. Current quoted turnaround is 21 days from receipt of samples. Questions can be sent to admin@novagenlaboratories.com.

Common questions

Why does one peptide show several peaks in a mass spectrum?

Each peptide molecule can carry a different number of protons when it leaves the electrospray. A molecule with three protons, one with four and one with five all reach the detector, and each appears at a different m/z. The peaks are the same compound in different charge states. Software converts the series back to one neutral mass.

What is a charge state in mass spectrometry?

A charge state, written z, is the number of charges an ion carries. For a peptide in positive ion mode it is usually the number of added protons. The instrument measures mass divided by charge, not mass alone, so the charge state must be known before the mass can be calculated.

How do you calculate the mass of a peptide from its m/z values?

Determine the charge state z, then calculate M = z x (m/z) minus z x 1.00728. The value 1.00728 is the mass of a proton. The charge state can be read from the spacing between isotope peaks, which is about 1.003 divided by z, or from the gap between neighbouring peaks in the charge series.

Why is electrospray ionisation called a soft ionisation technique?

It transfers little excess energy to the molecule. The peptide is carried into the gas phase inside evaporating droplets and acquires charge by protonation, so covalent bonds generally stay intact and the molecular ion is observed. Fragmentation can still occur if the source conditions are set harshly.

Does a matching mass prove a peptide is the right one?

No. A match confirms that a species of the expected mass is present. Two peptides with the same amino acids in a different order have the same mass. Mass is strong evidence of identity, but it does not establish sequence, and it does not measure purity.